| Product Features | ||
| Medical & ITE safety approvals | ||
| 2 MOPP input to output isolation | ||
| Low Leakage current≤100μA | ||
| DOE efficiency level VI | ||
| CoC V5 Tier 2(2016) | ||
| Standby power ≤0.075W | ||
| 5V outputs, up to 15W | ||
| Up to 5000m operating altitude | ||
| Interchangeable AC plugs | ||
| Protections against: short circuit/overload/ over voltage | ||
| Technical Data | ||
| Input Voltage:90~264VAC | ||
| Input Current:0.5A at 90VAC | ||
| Standby Power :0.075W | ||
| Environmental Requirements | ||
| Operationg Temperature: 0℃~45℃ | ||
| Storage Temperature:-20℃~60℃ | ||
| Operating Humidity:10%~90%RH,non-condensing | ||
| Storage Humidity::5%~90%RH | ||
| Operating Altitude:5,000 M | ||
| Safety Approvals | ||
| Safety aAgency/Mark | Medical | ITE |
| CB | IEC60601-1 | IEC60950-1 IEC62368-1 |
| UL | ANSI/AAMI ES60601-1 | UL62368-1 |
| CAN/CSA C22.2 NO.60601-1 | CAN/CSA C22.2 NO.62368-1 | |
| TUV Rheinland/Mark | EN60601-1 | |
| TUV Rheinland/GS | - | EN62368-1 EN60950-1 |
| CCC | - | GB4943.1 |
| CE | - | EN62368 |
| PSE | - | J62368 |
| FCC | - | FCC PART 15 |
| Mechanical Details | ||
| 101.5(L)*30(W)*57(H)mm | ||
| Applications: | ||
| Medical Device and ITE Equipment | ||
| Models & Ratings | ||||||||
Model No. | Voltage | Current | Ripple & Noise | Voltage Tolerance | Line& Load Regulation |
| Efficiency | Start Up delay |
UES18LCPU-XXXYYYSPA | 5.0 | 0.01-3.00 | 200 | ±6% | Line:±1% Load: ±5% | Y Y | 81.84% | ≤3s |